BioGuard Class II Type B1 Series Biological Safety Cabinets
Categories:
-
Biological Safety Cabinets
Gelman BioGuard Class II Type B1 Series Laminar Flow Biological Safety Cabinets are designed to provide a high degree of personnel, product and environment protection for work with low to moderate risk biological agents that require Biosafety Level 1, 2 or 3 containment.
- superior ergonomic design & styling
- airflow system with circulation to exhaust ratio of 30% : 70%
- negative pressure double-wall housings for containment
- all stainless steel work zone construction with radiused corners
- spacious work zone interior
- smooth & effortless sliding tempered glass window, closes completely
- microprocessor-controlled diagnostic system with digital display & audible alarm
- hourmeter
- delayed-automatic-off timer
- elapsed / event timer
- ISO Class 5 (Class 100) air cleanliness work zone for absolute product protection
- ZERO-LEAK airflow system for enhanced safety
- reliability-proven design & performance
- Gelman BioGuard Class II Type B1 Series Laminar Flow Biological Safety Cabinets are designed to provide a high degree of personnel, product and environment protection for work with low to moderate risk biological agents that require Biosafety Level 1, 2 or 3 containment.
- The BioGuard Class II Type B1 Series cabinets are used in applications where greater amounts of volatiles are used than are commonly allowed by a canopy connected Class Type A2 biological safety cabinet.
- The BioGuard Class II Type B1 Series cabinets maintain a minimum average inflow velocity of 100 fpm through the work access opening.
- An effective air barrier between the operator and the work zone, maintained by an inflow of room air into a full-width front grille in the work opening provides personnel and environment protection.
- The barrier air blends with approximately 30% of the contaminated work area air in a sump beneath the work surface and is partially recirculated back to the work area through a supply HEPA filter on top of the work zone.
- The recirculated air, composed largely of uncontaminated inflow air is HEPA-filtered to create a biological-clean work environment for product protection.
- The remaining contaminated work area air of approximately 70% is exhausted to the outside atmosphere via a dedicated facility exhaust system after HEPA filtration.
- The BioGuard Class II Type B1 Series cabinets have all biologically contaminated ducts and plenums under negative pressure.
- The GELMAN ZERO-LEAK airflow system ensures that all positive pressure zones are surrounded by negative pressure zones to contain potentially-hazardous aerosols.