Besaire Biological Safety Cabinets
Besaire BSC Class II Type B2 Total Exhaust Biological Safety Cabinets
Categories:
-
Besaire Biological Safety Cabinets
Gelman Besaire Class II Type B2 (Total Exhaust) Laminar Flow Biological Safety Cabinets are 100% Total Exhaust cabinets designed to provide a high degree of personnel, environment and product protection in applications involving low to moderate risk biological agents that require Biosafety Level 1, 2 or 3 containment.
- superior ergonomic design & styling
- glare-free stainless steel work zone construction with clear tempered glass side panels for maximum visibility
- spacious work zone interior
- smooth & effortless sliding tempered glass window, closes completely
- microprocessor-controlled diagnostic system with digital display & audible alarm
- hourmeter
- delayed-automatic-off timer
- elapsed / event timer
- ISO Class 5 (Class 100) air cleanliness work zone for absolute product protection
- ZERO-LEAK airflow system for enhanced safety
- reliability-proven design & performance
- Gelman Besaire Class II Type B2 (Total Exhaust) Laminar Flow Biological Safety Cabinets are 100% Total Exhaust cabinets designed to provide a high degree of personnel, environment and product protection in applications involving low to moderate risk biological agents that require Biosafety Level 1, 2 or 3 containment.
- The cabinets may be used for work with biological agents that have been treated with toxic chemicals and radionuclides required as an adjunct to microbiological studies.
- Gelman Besaire Class II Type B2 Series cabinets maintain a minimum average inflow velocity of 100 fpm through the work access opening.
- Ambient air is drawn in through a prefilter on top of the cabinet and is introduced to the work area through a supply HEPA filter that creates a biological-clean work environment for product protection.
- An effective air barrier between the operator and the work zone, maintained by an inflow of room air into a full-width front grille in the work opening provides personnel and environment protection.
- The barrier air blends with the contaminated work area air in a sump beneath the work surface and is totally exhausted to the outside atmosphere via a dedicated facility exhaust system after HEPA filtration, without recirculation in the cabinet or return to the laboratory.
- The GELMAN ZERO-LEAK airflow system ensures that all positive pressure zones are surrounded by negative pressure zones to contain potentially-hazardous aerosols.