Biological Safety Cabinets
RadGuard Class II Type A2 Biological Safety Cabinets For Compounding Radiopharmaceuticals
Categories:
-
Biological Safety Cabinets
Gelman RadGuard Class II Type A2 Biological Safety Cabinets are designed to provide a high degree of personnel, product and environment protection in applications involving the handling of radioactive materials and compounding of radiopharmaceuticals.
- superior ergonomic design & styling
- negative pressure double-wall housings for containment
- all stainless steel work zone construction with radiused & crevice-free corners
- complete lead shielding in the walls & floor of the work zone
- spacious work zone interior
- smooth & effortless sliding tempered glass window, closes completely
- microprocessor-controlled diagnostic system with digital display & audible alarm
- hourmeter
- delayed-automatic-off timer
- elapsed / event timer
- ISO Class 5 (Class 100) air cleanliness work zone for absolute product protection
- ZERO-LEAK airflow system for enhanced safety
- reliability-proven design & performance
- Gelman RadGuard Class II Type A2 Biological Safety Cabinets are designed to provide a high degree of personnel, product and environment protection in applications involving the handling of radioactive materials and compounding of radiopharmaceuticals. The cabinets are suitable for work with low to moderate risk biological agents that require Biosafety Level 1, 2 or 3 containment.
- The RadGuard Biological Safety Cabinet has complete lead shielding in the walls and floor of the work zone.
- The RadGuard Biological Safety Cabinets maintain a minimum average inflow velocity of 100 fpm through the work access opening.
- An effective air barrier between the operator and the work zone, maintained by an inflow of room air into a full-width front grille in the work opening provides personnel and environment protection.
- The barrier air blends with the contaminated work area air in a sump beneath the work surface and is partially recirculated back to the work area through a supply HEPA filter on top of the work zone.
- The recirculated air is HEPA-filtered to create an ultra-clean work environment for the preparation of drug products.
- The remaining contaminated work area air is exhausted to the laboratory or outside atmosphere after HEPA filtration.
- The RadGuard Biological Safety Cabinets have all contaminated ducts and plenums under negative pressure with an airflow circulation to exhaust ratio of 70% : 30%.
- The GELMAN ZERO-LEAK airflow system ensures that all positive pressure zones are surrounded by negative pressure zones to contain potentially-hazardous aerosols.